Diagnostic Devices, Inc. www.prodigymeter.com
Prodigy Talking Meter Owner’s Manual
20
21
Diagnostic Devices, Inc. 1.800.243.2636
Prodigy Talking Meter Owner’s Manual
Performing a Control Solution Test
Startwiththemeteroff.
STEP 1: Insert Test Strip
Insert a test strip with the contact bar end
entering into the test slot first. Push the test
strip as far as it will go without bending it. The
meter turns on automatically.
See Figure 18.
STEP 2: Mark as a Control Solution Test
After the “ ” symbol appears on display,
press the “M” button and a control solution
bottle symbol appears on display. With the
control solution bottle symbol on the the
display, the meter will not store your test
result in the memory. If you decide not to
perform a control solution test, press
the “M” button again and the control
solution bottle symbol will disappear.
See Figure 19.
Important: Be sure that you are in
ControlSolutionModesothatthetest
result will not be stored in the meter
memory.
Figure 19
Figure 18
Important Control Solution Test Information
•UseonlyProdigy
®
ControlSolutions.
•Checktheexpirationdateonthecontrolsolutionbottle.
Do not use if expired.
•ControlSolution,meterandteststripsshouldcometo
roomtemperature(68-77°F/20-25°C)beforetesting.
•Usewithinaperiodof90daysfromthedatethatyourst
openit.Recordthediscarddateonthecontrolsolution
bottle and discard after 90 days.
•Storethecontrolsolutiontightlyclosedattemperatures
below86°F(30°C).Donotrefrigerate.
Important: Prodigy
®
’sControlSolutionrangesare
located on the back of the test strip vial. They are not
recommended target ranges for your blood glucose level.